Reference Instrument
Precise instruments determine the dew point temperature by cooling a reflective surface until water vapour begins to condense upon it. A chilled mirror hygrometer provides a fundamental measurement because the temperature at which condensation forms relates directly to the vapour pressure of the air. This method is often used as a laboratory standard to calibrate secondary humidity sensors.
It relies on the laws of thermodynamics rather than the electrical properties of a material.
Optical Detection
An internal light source and a photodetector monitor the surface of the mirror to identify the exact moment of dew formation. As the chilled mirror hygrometer lowers the temperature of the reflective plate, the intensity of the reflected light drops when the first droplets appear. A feedback loop then maintains the mirror at this precise equilibrium temperature.
This stability allows for the continuous monitoring of the dew point in a controlled environment. Thermal sensors embedded just below the surface record the temperature with high accuracy. The instrument captures the exact point where the gas phase and liquid phase reach equilibrium.
Operational Limit
The presence of salts or oils on the mirror surface can shift the observed dew point and introduce a measurable bias. While a chilled mirror hygrometer is highly accurate, it requires frequent cleaning to maintain its status as a reference. High concentrations of chemical solvents in the air can also interfere with the cooling cycle.