Calibration Ratio
Analytical chemistry parameters define the ratio between the signal response of a detector and the concentration of a specific compound injected into the system. This metric, known as the response factor, is used to calibrate chromatography systems for quantitative analysis of migrating substances in packaging extracts. Establishing this ratio allows the instrument to convert a recorded peak area back into a precise mass concentration.
Detector Sensitivity
Flame ionization detectors show different sensitivities to different chemical classes, which means a single concentration does not yield a single peak size. Calculating a specific response factor for each target compound compensates for this uneven detector response. This calculation is especially important when measuring complex mixtures of mineral oils and plasticizers from recycled paper.
Standard Quantitation
Using an internal standard with a known response factor allows laboratories to estimate the concentration of compounds for which no pure reference standards are available. This approach is called semi-quantitative analysis, and it is widely used in screening tests for food-contact paperboard. By comparing the peak area of an unknown migrant to the peak area of the internal standard, the analyst can calculate the concentration using the relative response factor.
Accurate determination of this relative factor is necessary to avoid significant underestimation of migrating chemicals.