Analytical Error
Non-target volatile compounds generated by the thermal decomposition of sorbent materials or polymer substrates during heating skew the emission profiles of packaging materials. These thermal desorption artifacts can be misidentified as original solvents or migrants from the printed paperboard. By mimicking the retention times of actual contaminants, the generated species reduce the accuracy of gas chromatography measurements.
They complicate the interpretation of headspace analysis of printed films.
Degradation Mechanism
High temperature exposure during the desorption step causes polymer chain scission in the packaging material or oxidative degradation of the polydimethylsiloxane sorbents. The formation of thermal desorption artifacts increases when the desorber temperature exceeds the thermal limit of the analytical trapping column. Oxygen residues in the carrier gas stream accelerate this degradation process and produce benzaldehyde or cyclic siloxanes that are not present in the original sample.
This reaction is minimized by using high-purity helium and installing efficient oxygen traps in the gas supply lines.
Quality Control
Establishing a baseline using empty tube runs and blank tests defines the limits of reliable compound identification. When the artifact peak area exceeds five percent of the total chromatogram area, the analysis must be stopped and the sorbent tube repacked. This step ensures that reported emission levels represent true material volatility.